Install the RDE300i according to the following instructions from Section 5 of the RDE300i User Guide.
Step 1: Mounting
- Mount the RDE300i as shown in Section 6 of the RDE300i User Guide, observing all listed requirements for mounting location, use of mounting holes, and enclosure orientation.
Step 2: Power and Communication
- Install power and communication cables as per Section 7.1 of the RDE300i User Guide.
Step 3: M Cables
- Connect the RDE300i M+/M- cables to a single PV module, per Section 7.2 of the RDE300i User Guide and per the wiring diagram in Figure 7‑4.
WARNING: DO NOT INTERCHANGE MODULE AND STRING CABLES. Follow the wiring diagram carefully. Connections labelled “Module” or “M” must go to a single PV module. Do not interchange the Module and String inputs, which could result in serious damage to the product or interconnected equipment.
WARNING: DO NOT PROCEED UNTIL THIS STEP IS COMPLETE. Installing cables labelled “S” before cables labelled "M” may result in serious damage to the product or interconnected equipment.
Step 4: RTD
- Install the RTD on the Module, per Section 7.3 of the RDE300i User Guide .
Step 5: Disconnect String
For the 810275-01 RDE300i in-line version with String cables only:
- DISCONNECT THE STRING FROM ALL OTHER STRINGS, SOURCES, AND LOADS.
- OPEN THE IN-LINE FUSE THAT CONNECTS THE STRING TO THE PV STRING HARNESS, IF PRESENT.
- OPEN THE STRING CONNECTION AT THE STRING COMBINER BOX.
- DISCONNECT THE STRING FROM ANY OTHER CIRCUITS.
WARNING: THIS STEP MUST BE PERFORMED BY QUALIFIED PERSONNEL ONLY. FOLLOW ALL LOCAL AND SITE-SPECIFIC SAFETY REQUIREMENTS.
WARNING: DO NOT CONNECT OR DISCONNECT S-/S+ CABLES WHILE THE STRING IS CONNECTED TO ANY OTHER STRINGS, SOURCES, OR LOADS. During installation, the string must be disconnected from all other strings, batteries, power sources, loads, and inverters in the array. Failure to follow this instruction may result in damage to the product or interconnected equipment and/or bodily injury.
Step 6: S Cables
For the 810275-01 RDE300i in-line version with String cables only:
- Perform this step before or after daylight hours.
- Connect the RDE300i S-/S+ cables to complete the string circuit, per Section 7.2 of the RDE300i User Guide and per the wiring diagram in Figure 7‑4
WARNING: DO NOT CONNECT S-/S+ CABLES DURING DAYLIGHT HOURS. Connect the S cables before or after daylight hours to prevent possible damage to the product or interconnected equipment.
WARNING: DO NOT INTERCHANGE M+/M- AND S-/S+ CABLES. Follow the wiring diagram carefully. Connections labelled “Module” or “M” must go to a single PV module. Do not interchange the Module and String inputs, which could result in damage to the product or interconnected equipment.
WARNING: DO NOT CONNECT S-/S+ CABLES UNTIL STEP 3 – CONNECTION OF PV MODULE CABLES – IS COMPLETE. Failure to follow this instruction may result in damage to the product or interconnected equipment.
Step 7: Reconnect String
For the 810275-01 RDE300i in-line version with String cables only:
- Reconnect the in-line fuse to the PV string harness, if present.
- Reconnect the string connection at the string combiner box, if present.
Additional Steps
- For more information on system configurations that integrate the RDE300i as a Leader with other sensors including additional RDE300i units, the Mars™ soiling sensor, and/or RC22™ PV reference cell, see Atonometrics document 880100.
- For configuration of the RDE300i unit, see section 8 of the RDE300i User Guide .
Uninstalling
If it is necessary to uninstall the RDE300i:
- Disconnect the string per section 5.5
- Disconnect the S cables
- Disconnect the M cables
- Disconnect all other cables
Was this article helpful?
That’s Great!
Thank you for your feedback
Sorry! We couldn't be helpful
Thank you for your feedback
Feedback sent
We appreciate your effort and will try to fix the article